1. Mechanical
installation check
Check
whether there are mechanical installation flaws, whether the screws is less or
not tight, whether the circlip
is stuck, whether each cover is covered well.
2. Circuit
check
First
check whether the system is short circuit or open circuit, if everything is ok,
power test can be done, check whether the external power supply is stable,
whether transformer output voltage is too high or too low.
3. Balance
arm adjustment
Check
whether the balance arm and the lamp head are matching, whether the balance arm
force and angle should be adjusted.
4. Joint
flexibility adjustment
Mainly
adjust the joint damping screws, the standard stipulates that the degree of
tightness of the damping adjustment is close to 20N or 5Nm when push or rotate
the joint at any position.
5. Adjust
and open joint limit switch
In
order to facilitate the
surgery lamp installation and transportation, most
surgery lamp set lock switch or limit switch in its key components such as
joints, now open it.
6. Check
the irradiation distance and depth
Most
surgery lamp irradiation distance is between 70-140cm or 60-130cm, that means
surgery lamp works best in this distance.
7. Check
illuminance and color temperature
Illuminance
and color temperature are described for surgery
lamp irradiation intensity. Usually the best
surgery light is similar to sunlight, that is around 4300-4600K.
After those inspection the surgery lamp
can be put into use.