Medical
air purification disinfection provides an important guarantee for hospital
operating room, maternity ward, burn ward and intensive care unit’s air
quality, more frequently usage, so how to maintain Mingtai
medical air purification disinfection in
daily usage?
1.
When air purification disinfection
works, objects or hands are prohibited to closing to equipment import and
export ventilator, good protection measures should be done when handling to
avoid hard objects impact or fell to the ground.
2.
Once the warning symbol appears, stop
using immediately, turn off the power switch, pull out the plug and ask
relevant personnel to check, and read the product description carefully to
identify the cause of the alarm, after solving it then use it.
3.
Every 3 months check and clean the
filter: unplug the power outlet, open the air into panel to remove the filter, wash
with water, it is strictly prohibited with brush tools, water temperature
should not exceed 40 degree, so as not to deformation, washing and placing cold
and ventilated place, installed according to the original location, every year
to replace the filter. Cleaning, replacement filters should be recorded.
4.
Every 6-12 months check and clean dust
inside the machine: use soft brush, cloth or alcohol to wipe the surface of
each part of the machine. When cleaning the electrostatic adsorption unit,
clean it after power off 5 minutes. Cannot use gasoline, acid and alkali
solution cleaning machine.
5.
When usage time reach more than 2000
hours, ask relevant personnel to check the UV lamp to determine whether to
replace. Replace records should be recorded.
6.
Opening the lid when using air
purification sterilizer is strictly prohibited to prevent damage to people and
machines. Maintenance should also pay attention to their own safety.
7.
Do not cover the top of the equipment.
8.
Keep the equipment clean and dry. Power should
be off when clean the surface of equipment, to prevent the infusion accidents.