Electric operating
table is very popular now, and it utilization rate in major hospitals is
also very high, although operating table design life is very long, however, due
to improper operation during normal use, or without regular maintenance, over
time there will be some problems, such as the operating table will incline when
use it, or automatic fall. Shandong Mingtai
Medical Equipment Group Co., Ltd glad to tell you how to resolve it.
Firstly
determine whether the solenoid valve failure, 2 ways: 1. use three-meter
measuring resistance; 2. put metal on it to see whether there is suction. If
the solenoid valve pull action is normal, the obstruction of the oil will also cause
above problems.
Secondly
determine whether the compression pump failure. First check whether there is
voltage on compression pump, use three-meter measuring resistance, if the above
are normal is generally commutation capacitor failure.
There
is action in one side, no action in another side. Unilateral non-operational
failure is generally caused by electromagnetic reversing valve. Electromagnetic
reversing valve fault may be caused by bad control circuit, also may be
reversing valve mechanical stuck. The specific inspection method is check
whether there is voltage on reversing valve, if there is voltage then open reversing
valve to clean it. Due to long-term use, if there is a little impurity on the
movable shaft of the valve, the shaft will be burnt, which will cause the
operation table move in only one direction.
Operating
table will be automatically lowered down slowly. This situation is much more occurred
in mechanical operating table, mainly because of pump failure. Small impurities
may remain in the inlet valve after few years’ usage, resulting in minor
leakage. The solution is to open the lift pump and clean it with gasoline, in
particular, to check the inlet valve.