1. No gas source output from medical gas
Common causes
Common causes
(1). Whether the maintenance valve is open
(2). Medical gas pipeline is not connected or the air supply
system is not aspirated
(3). The secondary terminal is not inserted in correct place
Solution
(3). The secondary terminal is not inserted in correct place
Solution
(1). Open the maintenance valve
(2). Connect medical gas pipeline or the air supply system support
gas
(3). Insert the secondary terminal in correct place
(3). Insert the secondary terminal in correct place
2. Medical gas terminal leak
Common causes
(1). Medical gas terminal is not reset
(2). The terminal seal is damaged
(3). There are debris in the gas source to catch the terminal
Solution
(3). There are debris in the gas source to catch the terminal
Solution
(1). Use the secondary terminal plug repeatedly plug and pull
to reset it
(2). Change medical gas terminal
(3). Clear the debris in the gas channel and terminal
(3). Clear the debris in the gas channel and terminal
3. Power indicator does not light/ Power outlet no power
Common causes
(1). The power cord is not connected
(2). The power circuit breaker is not dialed
Solution
Solution
(1). Connect power cord
(2). Dial power circuit breaker
4. Button does not work, work indicator light does not light
or light but the brake and lift does not move
Common causes
Common causes
(1). The switching power supply is damaged
(2). The instrument platform overloads makes the overload
protection circuit work
(3). Motor short circuit makes overload protection circuit work
(3). Motor short circuit makes overload protection circuit work
(4). Poor connector contact or control line short circuit
makes no control signal to the electrical box input, causing no work
(5). Circuit board damaged or poor component contact and weld
Solution
Solution
(1). Change switching power supply
(2). Remove the heavy objects on the instrument platform
(3). Repair or replace the motor
(3). Repair or replace the motor
(4). Dispose or replace the connector, control line
(5). Replace the circuit board
5. Network or telephone terminal no signal output
Common causes
(1). Network output module has bad contact with network
cable, crystal head or wiring error; network short circuit
(2). Telephone output module has bad contact with network
cable, crystal head or wiring error; telephone line short circuit
Solution
(1). Reconnect the network output module and the crystal head,
replace the network cable
(2). Reconnect the telephone output module and the crystal
head, replace the telephone line
6. Arm and terminal box cannot be rotated or cannot be
positioned, the device drift phenomenon
Common causes
(1). Spherical nuts are tightly stuck
(2). Brake device is damaged or power supply, air supply is
not connected
(3). Brake positioning device contact is not good
Solution
(1). Adjust the damping brake lock nut
(2). Connect the power supply and air supply or replace the
brake
(3). Adjust the brake positioning device
7. Medical pendant shake when lift, instrument platform tilt,
the machine is tilted
Common causes
(1). The pipe upper or lower connecting screw is loose, the
platform connecting screw is loose
(2). Medical pendant assembly and bracket connection loose or
bracket and concrete floor connection loose
Solution
(1). Adjust the fastening screw
(2). Fastening bracket connection bolts
8. Arm, terminal box and instrument platform rotation more
than 340 °, cannot be limited
Common causes
(1). Locating pin broken because of excessive force when
rotation
(2). Locating pin off
Solution
(1). Change locating pin
9. Arm, terminal box and instrument platform rotation is not flexible
Common causes
(1). The upper part of the pipe is tightly pressed
(2). The center beam of the middle beam is too tight
Solution
(1). Adjust the tightening nuts